March 08, 2025 11:36 PM
Siemens Healthineers has confirmed, through the investigation of customer complaints, the presence of sodium hypochlorite (NaOCl) contamination in certain bottles of Atellica CH Diluent lot number 0000191602, bottle sequence numbers 00001-04400 (cartons 00001-02200) used on Atellica CH 930 and Atellica CI analyzers. Atellica IM assays are not affected. As a result of this issue, calibrator, quality control (QC), and patient results could be affected for any Atellica CH assay excluding sodium, potassium, and chloride. This issue is specific to Atellica CH Diluent lot number 0000191602. Unaffected alternate lots of this product are available.
Starting Date: February 26, 2025
Description:
Siemens Healthineers has identified sodium hypochlorite (NaOCl) contamination in certain bottles of Atellica CH Diluent, specifically lot number 0000191602, bottle sequence numbers 00001-04400 (cartons 00001-02200). This issue may impact calibrator, quality control (QC), and patient results for Atellica CH assays, excluding sodium, potassium, and chloride tests. Atellica IM assays are not affected. Siemens Healthineers has confirmed that unaffected alternative lots are available for use.
For more information on this recall, see the: Health product recall – Atellica CH Diluent
The information provided on the recall section of this website is for general awareness only. While we strive for accuracy, users should verify details with official authorities. We are not responsible for any actions taken based on this information. See Disclaimer Policy for more details.